SIGLENT SDS1000D Manual de usuario Pagina 110

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 154
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 109
102
2.13.4 Self Calibration
Self Calibration procedure can optimize the signal path at the most measurement
precision..You can run this procedure at any time. If the operating temperature changes
by more than 5° C or the units run more than thirty minutes,you should do the self
calibration.
When you run the self Calibration, you should cut off all the probes and lead. Then
press the “Utility” button to choose the “Do self cal” to show the self Calibration
Menu, and execute self calibration program according to the prompting on the
screen.
Picture 2.13-10
2.13.5 Self Test
Press “UTILITY” “Do Self Test”
Table2-55
Operation Steps:
1. Screen Test:
Select “Screen Test” to enter the screen test interface. The clew words “Press ‘SINGLE’
Key to continuePress ‘RUN/STOP’ Key to exit” is displayed, you could press the
Option Introduction
Screen Test Run Screen Test Program.
Keyboard Test
Run Keyboard Test Program.
LED Test Run LED Test Program.
Vista de pagina 109
1 2 ... 105 106 107 108 109 110 111 112 113 114 115 ... 153 154

Comentarios a estos manuales

Sin comentarios