85
Picture 2.12-7
Save setup to USB flash drive, follow these steps
For example: Save setup that set waveform display type to “Dots” to USB flach drive.
Press the “SAVE/RECALL” button to select “Setups”.
1 Insert USB flash drive to USB host port of the oscilloscope and wait that
the oscilloscope has initialized USB flash drive (about five second).
2 Press the “Save to” option button to select “file”.
3 Press the “Save” option button then you’ll go into the Save/Recall
interface.
4 Press the “New folder” button to creat a new folder
5 Press the “Del folder” option button to delete a folder
6 Press the “Modify” option button to modify a folder.
7 Press the “New file” option button to creat a new file
8 Press the “Del file” option button to delete a file
10) Press “the next page ” option button and press the “Rename” option button
to modify the name of the file or the folder.
11) Turn the “Universal” knob to select the folder and press the
“Confirm”option button and you save the set to the USB drive.